20
Password-protected programs only:
Message
If the operator chooses to attempt to start the sterilizer, even though
starting is blocked, the operator panel display shows a message
referring to an essential condition that is not met.
JACKET TEMP HIGH - Indicates that lowering of jacket
temperature is in progress.
JACKET TEMP LOW - Indicates that raising of jacket temperature
is in progress.
STEPPING KEY ON - Indicates that key switch for stepping is on.
PRINTING ACTIVE - Indicates that printing from the printer is
blocking the start of a new program.
UNACKNOWLEDGED ALARM – Indicates that a previous alarm
has not been acknowledged and blocks the start of a new program.
SEAL PRESSURE – Indicates that the door seal is pressurised.
DOOR KEYSWITCH - Indicates that the keyswitch for cleaning is
on.
DOOR NOT CLOSED - Indicates that a sterilizer door is not
properly closed.
LOW WATER LEVEL – Indicates that the water level in the tank is
low.
COMPRESSED AIR – Indicates that the pressure of the air supply is
low.
DEGASSING TANK – Indicates that the temperature in the
degassing tank is low.
START CONDITION - Indicates that a starting condition specific to
the type of unit is not met.
(Steam, cooling water, formalin bottle or ventilation at formalin
sterilizers)
Alarms
If a fault occurs, in standby or during a process, the system goes to an
alarm phase. This is an exceptional situation, but it is normally not
dangerous. The current process is stopped and output signals from the
control system go to settings that maintain the safety of personnel, the
unit and the load.
After the start button has been pressed, the operator is
prompted to enter the password.
[START]