21
The events below take place when the unit goes into an alarm phase.
The alarm is registered automatically in the control system fault log-.
The fault log- contains information about the last 20 alarms.
The fault codes are explained later in this chapter.
Blowing safety valve
If a safety valve blows off:
Press the EMERGENCY STOP button.
If this has no effect, cut off the electric power supply with the control
switch on the front of the unit.
In an emergency, the power supply can be cut off with the working
switch.
Send for a service technician without delay.
An audible signal sounds.
The process fault alarm indicator
flashes.
/
The type of fault appears on the unit display in plain text.
If the unit has a printer, the alarm text is printed out as part of the
process log.
If an alarm occurs:
Read the message on the panel.
In case of an error concerning a temperature sensor, a
pressure transducer or the Back-up battery a service
technician is to be called for.
Press button to acknoledge the
alarm, The audible signal stops and
the indication changes from
flashing to a steady light
/ [ALARM OK]