22
Alarms in standby mode
An alarm from an idle sterilizer prevents it being started.
• Call a technician.
Alarms during a process
A fault occurring during a process stops the program. All valves close,
stabilising the conditions in the chamber..
On power failure or if the control switch or working switch
of the sterilizer had to be turned off during a process, there
is a risk of leakage at the door. Safety measure when there
is positive pressure in the chamber:
KEEP WELL AWAY FROM THE DOOR AND CALL A
TECHNICIAN!
If the load contains liquids, choosing the wrong program
may endanger personnel and equipment.
The process complete lamp lights up red.
There are two options:
• Depending on how the program is set up, the process is
completed automatically, or the process stops until you
press the button. The process is then completed auto-
matically.
• Call a technician, who will operate the Stepping keyswitch to start
the process so that it continues from where it stopped. The fault
code remains on the display.
For safety reasons, it is not possible to step past a
pressure equalisation phase or a cooling phase.