4
Maintenance and inspection procedures
Mechanical inspections
22 / 44
Lucea 50_100
01740 EN 01
Fig.21: Checking the caps
Check for any loose covers and caps.
Fig.22: Stability/drift
Stability and drift of the system
1. Operate the device, making several move-
ments in order to swivel the extension
arms, the spring arms and the lightheads.
Ø The entire system should move easily
and smoothly.
2. Place the system in various positions.
Ø The entire system should remain in the
selected position, without any drift.
3. If a problem is noted, contact technical
support.
Fig.23: Checking the balance
Check the balance of the spring arm and
limit stops.