SERVICE MANUAL ProSeal Classic/Plus/Premium
31
Press the + button to turn on the motor forward. On the right in the display you
can read the speed of the running motor; the value should be the same (± 10
mm/sec).
For validation adjust the speed to 120 mm/s (min. speed); check with a
tachometer if the speed of the belt is equal, check also at 170 mm/s
(max. speed).
The factory setting for speed is 160mm/s.
5.3.4 Temperature
Be aware of the tolerance of the temperature device you use.
Pic 50
Go back to the validation menu and press Temp.
In this Temperature menu (Pic 50) you see the temperature set point and the
measured temperature of the unit. There will always be a difference between
these values which is caused by the temperature control.
Place heat conduction paste on top of the heat element (Pic 51).
Pic 51
Change the set temperature to 100 °C
Place the temperature probe on top of the heat conducting paste.
Compare the temperature of the upper heat element with the temperature
meter. Be sure that the measured temperature is the same as the temperature
of the upper heat element (± 5°C).
Change the set temperature to 200 °C
Compare the temperature of the upper heat element with the temperature
meter. Be sure that the measured temperature is the same as the temperature
of the upper heat element (± 5°C).
6 Spare parts
Spare parts can be ordered by means of the enclosed drawings. Always state the
type designation and the serial number of the unit when ordering spare parts. Copy
the respective drawing and mark the required spare part.