Ultrasonic Anemometer
WindObserver Sensor and Bracket Mounting
Mount the Sensor on to the Bracket so that the Sensor North Marker points to the bracket pole.
Secure the sensor to its mount using the three M5 screws and three M5 split wavy washers supplied in the mounting
bracket kit box. Torque the three M5 bolts and washers to 4.5Nm.
The Bracket uses a mounting clamp suitable for attaching to a vertical pipe with a diameter of 30-58mm. When
mounting the Bracket Sensor, consider the position, orientation and alignment of the unit. Mount at the top of a Pipe to
ensure a clear unobstructed measurement view.
Note that the mounting pipe should first be degreased and when assembling the Bracket clamp assembly, the outer
clamp nuts need to be tightened evenly to a torque figure of 3Nm.
The moving plate part of the clamp needs to be reversed for poles below 38mm diameter.
Screw an Earth cable minimum of 6mm
2
to the Bracket chassis using Screw and Terminal Tag fixings supplied.