Ultrasonic Anemometer
The polled mode commands available in this mode are:
Description Command WindObserver response
Enable Polled Mode ? (None)
WindObserver Unit Identifier (N) A…….Z Output generated
Disable Polled mode ! (None)
Request WindObserver Unit Identifier & A ….. Z (as configured)
Enter Configuration mode *<N> CONFIGURATION MODE
Where <N> is the unit identifier, if used in a multi-drop system then it is recommended that ID’s A to F and KMNP are not
used as these characters can be present in the data string.
It is suggested that in polled mode the following sequence is used for every poll for information.
? Ensures that the Sensor is enabled to cover the event that a power down has occurred.
A-Z Appropriate unit designator (capitol letter) sent to retrieve a line of data.
! Sent to disable poll mode and reduce possibility of erroneous poll generation.
When polled with the default Mode 14 factory setting the WindObserver (set for default 1Hz output) will output the last
valid 10-minute wind speed and direction average, updated every minute along with last valid 1 minute Gust
magnitude.
Non Heat Enabled Units.
If the unit is powered up and polled before the unit has reached its minimum averaging interval the status code will read
51 (Measurement Average Building, non-heat enabled units only).
Heat Enabled Units.
On units where heat is enabled Code 51 will not be output if polled during the measurement average building period,
therefore allow for this period on power up before taking readings to be fully averaged (maximum period of 10 minutes).