51
Issue: 8 (Applies to firmware 2387 6.09 onwards)
Document number: 1390-PS-0039
Returning to Measurement mode
Type Q and press ENTER
If in Continuous mode, the anemometer responds with wind measurements immediately, continuing at the selected
Sampling rate.
If in Polled mode,
? Enables poll
N Polls anemometer (where N is the Unit identifier entered as upper-case)
The anemometer replies with a single set of wind measurements
& Anemometer replies with Unit identifier
! Disables poll
Note:- If the unit is in Polled Mode it is always best to interrogate the unit for data with a ? Before the poll identifier to
cater for the event that the power has been switched off or power interrupted.
If used in a multi drop system then it is recommended that Unit Identifiers A, B, C, D, E, F, K, M, N and P are not used.
Checking the configuration
We strongly recommend that, as a standard procedure, you use this command ( D3 ) prior to, and after, changing any
settings. It shows the current settings for all the alterable settings. We suggest you note down your settings, so that you
can easily return to them.
Type
* Enters Configuration Mode (from Continuous mode)
Or Type *N Enters Configuration Mode (from Polled mode)
T
ype D3 and press ENTER The WindObserver 65 responds with the current configuration settings.
The factory default settings for a heated unit are: -
A
0 B3 C1 E1 F1 G0000 H2 J1 K1 L1 M2 NA O1 P1 T1 U1 V1 X1 Y1 Z1
Or factory default settings for an unheated Unit are:-
A
0 B3 C1 E1 F1 G0000 J1 K1 L1 M2 NA O1 P1 T1 U1 V1 X1 Y1 Z1
To return to Measurement mode Type Q and press ENTER How to change these settings is explained in the
following sections.