EasyManuals Logo

Gima RS01 User Manual

Gima RS01
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
41
Appendix 2
Guidance and manufacturer’s declaration – electromagnetic emissions – for
all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity – for
all EQUIPMENT and SYSTEMS
ENGLISH
Guidance and manufacturer’s declaration – electromagnetic emission
The RS01 is intended for use in the electromagnetic environment specied below. The
customer of the user of the RS01 should assure that it is used in such and environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The RS01 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The RS01 is suitable for use in all establishments
other than domestic and those directly connected
to a low voltage power supply network which sup-
plies buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
The RS01 is intended for use in the electromagnetic environment specied below. The
customer or the user of RS01 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If oor are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Power frequency
(50/60Hz)
magnetic eld
IEC 61000-4-8
3A/m Class B Mains power quality should be
that of a typical commercial or
hospital environment.

Other manuals for Gima RS01

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gima RS01 and is the answer not in the manual?

Gima RS01 Specifications

General IconGeneral
BrandGima
ModelRS01
CategoryMedical Equipment
LanguageEnglish

Related product manuals