42
Guidance and manufacturer’s declaration – electromagnetic immunity – for
EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
80 MHz - 800 MHz
800 MHz - 2,5 GHz
d =
[
]
√ P
3.5
E
1
d =
[
]
√ P
7
E
1
d =
[
]
√ P
3.5
V
1
ENGLISH
Guidance and manufacturer’s declaration – electromagnetic immunity
The RS01 is intended for use in the electromagnetic environment specied below. The
customer or the user of RS01 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the RS01, includ-
ing cables, than the recommended
separation distance calculated from
the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
Where P is the maximum output pow-
er rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended sep-
aration distance in metres (m).
Field strengths from xed RF trans-
mitters, as determined by an electro-
magnetic site survey,
a
should be less
than the compliance level in each fre-
quency range.
b
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reection from structures, objects and people.