46 47
FULL-ON AND REFERENCE TEST GAIN
MAXIMUM OUTPUT (OSPL 90)
Note: Telecoil sensivity only applicable for 62 RIE models
Reference test gain (60 dB SPL input) HFA 37 dB
Full-on gain (50 dB SPL Input)
Max
HFA
58
51
dB
dB
Maximum output (90 dB SPL input)
Max
HFA
116
114
dB SPL
dB SPL
Total harmonic distoron
500 Hz
800 Hz
1600 Hz
0.5
0.6
1.2
%
%
%
Telecoil sensivity (SPLIV @ 31.6 mA/m) 96 dB SPL
Equivalent input noise (w/o noise reducon) 23 dB SPL
Frequency range (DIN 45605) 100–7000 Hz
Current drain (in test mode) 1.3 mA
RIE—MP receiver
AM561-DRW, AM562-DRW
Frequency (Hz)
Gain
(dB)
Reference test gain
60 dB SPL input
Full-on gain
50 dB SPL input
2cc Coupler
Frequency (Hz)
Output
(dB SPL)
2cc Coupler
21 Technical data
Reference test gain (60 dB SPL input) HFA 42 dB
Full-on gain (50 dB SPL Input)
Max
HFA
65
56
dB
dB
Maximum output (90 dB SPL input)
Max
HFA
122
118
dB SPL
dB SPL
Total harmonic distoron
500 Hz
800 Hz
1600 Hz
0.6
1.2
0.7
%
%
%
Telecoil sensivity (SPLIV @ 31.6 mA/m) 101 dB SPL
Equivalent input noise (w/o noise reducon) 23 dB SPL
Frequency range (DIN 45605) 100–6030 Hz
Current drain (in test mode) 1.3 mA
RIE—HP receiver
AM561-DRW, AM562-DRW
Frequency (Hz)
Gain
(dB)
Reference test gain
60 dB SPL input
Full-on gain
50 dB SPL input
2cc Coupler
Frequency (Hz)
Output
(dB SPL)
2cc Coupler
FULL-ON AND REFERENCE TEST GAIN
MAXIMUM OUTPUT (OSPL 90)
Note: Telecoil sensivity only applicable for 62 RIE models
Data in accordance with IEC60118-7 and
ANSI S3.22-2009, supply Voltage 1.3V
Data in accordance with IEC60118-7 and
ANSI S3.22-2009, supply Voltage 1.3V