Initial setup steps
Installation
WARNING –
Potential electrostatic hazard
Removing protective packaging and
lm may cause electrostatic discharge.
Protective packaging and lm shall not
be removed in ATEX zones.
I-1*: Due to safety and operating considerations, the cabinet must not be used outdoors.
I-2*: The cabinet should be installed in a dry and suciently ventilated area.
I-3*: To ensure ecient operation, the cabinet should not be installed in direct sunlight or close to heat sources.
The cabinet interior must not be exposed to corrosive atmospheres.
I-6*: The cabinet is shipped with a protective lm that should be removed prior to use.
I-5*: Avoid placement of the cabinet in a chloric/acidic environment due to risk of corrosion.
I-7*: Clean the cabinet with a mild soap solution prior to use.
I-8*: The cabinet is only allowed to lay down for very short durations (for instance handling through a
doorway). If the cabinet has been laying down, the cabinet must stand up-right for at least 24 hours
prior to use. This enables oil in the compressors to run back into place.
*): Paragraph designation (I-1, I-2 etc) refers to the IQ (Installation Qualication)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram bioline BioCompact II 210 and is the answer not in the manual?

Gram bioline BioCompact II 210 Specifications

General IconGeneral
BrandGram
Modelbioline BioCompact II 210
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals