31
b i o l ineb i o l ine
O-6*: LLd – Setting the delay of the local lower alarm limit [min.]
Press and hold + for more than 3 seconds
Press to select “LAL”. “LhL” is now shown in the display
Press several times until “LLd” is shown in the display
Press to select “LLd”. The delay of the lower alarm limit is now shown in the display
Press or to set the desired value for the delay of the lower alarm limit
Press to conrm the set value
– The delay of the lower alarm limit is now set, proceed to other parameters by pressing , and then navigate by
using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
O-5*: Lhd – Setting the delay of the local upper alarm limit [min.]
Press and hold + for more than 3 seconds
Press to select “LAL”. “LhL” is now shown in the display
Press several times until “Lhd” is shown in the display
Press to select “Lhd”. The delay of the upper alarm limit is now shown in the display
Press or to set the desired value for the delay of the upper alarm limit
Press to conrm the set value
– The delay of the upper alarm limit is now set, proceed to other parameters by pressing , and then navigate by
using or .
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
Local alarm settings
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum
item safety.
*): Paragraph designation (O-1, O-2 etc) refers to the OQ (Operation Qualication)