EasyManuals Logo

Gram bioline BioCompact II 210 User Manual

Gram bioline BioCompact II 210
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
59
b i o l ine
BioCompact II RF410H
Technical specications Data
Temperature range -25/-5 °C
Ambient temperature range +10/+35 °C
Software variant F52
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T3 Gc
ATEX certicate DTI 22ATEX0251X
Gross volume 346 litres
Net volume 312 litres
Dimensions – W x D x H 595 x 640 x 1776/1976 mm
Refrigerant R134a
Refrigerant charge 95 g
Refrigeration capacity at -25 °C 216 Watt
GWP – CO2e 135.85
Energy consumption –
default setpoint
2.66 kWh/24h
Heat emission 100 % 190 Watt
Heat emission default setpoint 111 Watt
Nominal consumption 191 Watt
Start current 11.3 A
Sound level 38.6 dB(A)
BioCompact II RF410H
Solid door

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram bioline BioCompact II 210 and is the answer not in the manual?

Gram bioline BioCompact II 210 Specifications

General IconGeneral
BrandGram
Modelbioline BioCompact II 210
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals