56
b i o l ine
BioCompact II RR310H
Technical specications Data
Temperature range +2/+20 °C
Ambient temperature range +10/+32 °C
Software variant K92
K-Value 0.50 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T5 Gc
ATEX certicate DTI 22ATEX0251X
Gross volume 218 litres
Net volume 189 litres
Dimensions – W x D x H 595 x 640 x 1190/1390 mm
Refrigerant R134a
Refrigerant charge 95 g
Refrigeration capacity at -10 °C 173 Watt
GWP – CO2e 135.85
Energy consumption –
default setpoint
1.30 kWh/24h
Heat emission 100 % 180 Watt
Heat emission default setpoint 56 Watt
Nominal consumption 183 Watt
Start current 9.7A
Sound level 36.8 dB(A)
BioCompact II RR310H
Glass door
Technical specications Data
Temperature range +2/+20 °C
Ambient temperature range +10/+35 °C
Software variant K82
K-Value 0.35 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T5 Gc
ATEX certicate DTI 22ATEX0251X
Gross volume 218 litres
Net volume 189 litres
Dimensions – W x D x H 595 x 640 x 1190/1390 mm
Refrigerant R134a
Refrigerant charge 115 g
Refrigeration capacity at -10 °C 173 Watt
GWP – CO2e 164.45
Energy consumption –
default setpoint
0.79 kWh/24h
Heat emission 100 % 145 Watt
Heat emission default setpoint 33 Watt
Nominal consumption 145 Watt
Start current 8.6A
Sound level 36.8 dB(A)
BioCompact II RR310H
Solid door