ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum
item safety.
External acoustic settings
O-16*: BU – Activation/deactivation of the acoustic external alarms
Press and hold + for more than 3 seconds
Press to proceed to “EAL”
Press to select “EAL”. “EhL” is now shown in the display
Press several times until “BU” is shown in the display
Press to select “BU”
Press or to activate/deactivate the external acoustic alarms [1 = activated/0 = deactivated]
Press to conrm the set value
– The external acoustic alarms is congured, proceed to other parameters by pressing , and then navigate by
using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
*): Paragraph designation (O-1, O-2 etc) refers to the OQ (Operation Qualication)