Technical specications Data
Temperature range +2/+20 °C
Ambient temperature range +10/+35 °C
Software variant K82
K-Value 0.35 W/(m2*K);
ATEX marking II 3G Ex ec nC ic llB T3 Gc
ATEX certicate DTI 22ATEX0251X
Gross volume 2 x 125 litres
Net volume 2 x 104 litres
Dimensions – W x D x H 595 x 640 x 1602/1802 mm
Refrigerant R134a
Refrigerant charge 160 g
Refrigeration capacity at -10 °C 150 Watt
GWP – CO2e 228.8
Energy consumption –
default setpoint
1.36 kWh/24h
Heat emission 100 % 222 Watt
Heat emission default setpoint 56 Watt
Nominal consumption 220 Watt
Start current 13.8 A
Sound level 39.1 dB(A)
BioCompact II RR210/RR210H
Solid door
BioCompact II RR210/RR210H
Technical specications Data
Temperature range +2/+20 °C
Ambient temperature range +10/+32 °C
Software variant K92
K-Value 0.49 W/(m2*K)
ATEX marking II 3G Ex ec nC ic llB T3 Gc
ATEX certicate DTI 22ATEX0251X
Gross volume 2 x 125 litres
Net volume 2 x 104 litres
Dimensions – W x D x H 595 x 640 x 1602/1802 mm
Refrigerant R134a
Refrigerant charge 160 g
Refrigeration capacity at -10 °C 150 Watt
GWP – CO2e 228.8
Energy consumption –
default setpoint
2.44 kWh/24h
Heat emission 100 % 376 Watt
Heat emission default setpoint 102 Watt
Nominal consumption 376 Watt
Start current 13.8 A
Sound level 39.1 dB(A)
BioCompact II RR210/RR210H
Glass door

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram bioline BioCompact II 210 and is the answer not in the manual?

Gram bioline BioCompact II 210 Specifications

General IconGeneral
BrandGram
Modelbioline BioCompact II 210
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals