b i o l ine
O-17*: dEF – Number of defrosts
Press and hold + for more than 3 seconds
Press several times until “dEF” is shown in the display
Press to select “dEF”
Press or to set the desired amount of defrosts per 24 hours (factory setting is 4)
Press to conrm the set value
Leave the user menu by pressing several times until the cabinet temperature is shown in the display.
Defrosts/24 hours
Please note
It is very important that defrosts should not
be set to 0 for a prolonged period of time,
as this will reduce the cooling capacity of the
cabinet.
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum
item safety.
*): Paragraph designation (O-1, O-2 etc) refers to the OQ (Operation Qualication)