On/o local door alarm
Delay for local door alarm
O-8*: dAd – Setting the delay of the local door alarm [min.]
Press and hold + for more than 3 seconds
Press to select “LAL”. “LhL” is now shown in the display
Press several times until “dAd” is shown in the display
Press to select “dAd”. The delay of the local door alarm is now shown in the display
Press or to set the desired value for the delay of the local door alarm
Press to conrm the set value
– The delay of the local door alarm is now congured, proceed to other parameters by pressing , and then navigate
by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
O-7*: dA – Activate/deactivate of local door alarm
Press and hold + for more than 3 seconds
Press to select “LAL”. “LhL” is now shown in the display
Press several times until “dA” is shown in the display
Press to select “dA”.
Press or to activate/deactivate the local door alarm [1 = activated/0 = deactivated]
Press to conrm the set value
– The local door alarm is now congured, proceed to other parameters by pressing , and then navigate by using or
.
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum
item safety.
*): Paragraph designation (O-1, O-2 etc) refers to the OQ (Operation Qualication)