Display sensor
O-18*: dPS – Display sensor
Press and hold +
for more than 3 seconds
Press
several times until “dPS” is shown in the display
Press to select “dPS”
Press or to select either the A- or E-sensor
Press to conrm the set value
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
Please note: that dPS only changes the reference sensor for the display, and not the reference sensor for the alarms.
Please note: that the reference sensor for the refrigeration system is the A-sensor, this can not be altered.
ATTENTION –
High and low temperature alarms set
up in the cabinets’ controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum
item safety.
FP – Activation/deactivation of low temperature protection
Press and hold + for more than 3 seconds
Press several times until “FP” is shown in the display
Press to select “FP”. “Act” is now shown in the display
Press to select “Act”
Press or to activate/deactivate [1 = activated/ 0 = deactivated]
Press to conrm the set value
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
FP – Setpoint of low temperature protection
Press and hold + for more than 3 seconds
Press
several times until “FP” is shown in the display
Press
to select “FP”. “Act” is now shown in the display
Press
several times until “SEt” is shown in the display
Press
to select “SEt”
Press or to select the setpoint temperature for the low temperature protection
Press to conrm the set value
Leave the user menu by pressing several times until
the cabinet temperature is shown in the display
FP – Test of low temperature protection
Press and hold + for more than 3 seconds
Press several times until “FP” is shown in the display
Press to select “FP”. “Act” is now shown in the display
Press to progress to “tES”
Press to select “tES” – test will then be performed
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
FP – Temperature of low temperature protection sensor
Press and hold + for more than 3 seconds
Press several times until “FP” is shown in the display
Press to select “FP”. “Act” is now shown in the display
Press several times until “Pre” is shown in the display
Press to select “Pre”
Press to show the low temperature protection sensor temperature
Leave the user menu by pressing several times until the cabinet temperature is shown in the display
Electric low temperature protection
*): Paragraph designation (O-1, O-2 etc) refers to the OQ (Operation Qualication)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gram bioline BioCompact II and is the answer not in the manual?

Gram bioline BioCompact II Specifications

General IconGeneral
BrandGram
Modelbioline BioCompact II
CategoryMedical Equipment
LanguageEnglish

Related product manuals