93
b i o l ine
92
IQ & OQ
Installation Qualication
Operation Qualication
The following IQ / OQ is intended
to be a guideline, local IQ / OQ
procedures can vary depending on
application and items stored in the
Gram BioLine cabinet.
Deviations from the specications
dictated in the PQ are to be
reported in the deviation report.
The IQ / OQ is concluded if all
criteria of acceptance are approved
and the possible deviations are
rectied or accepted.
Organisation:
Location of installation:
Model:
Serial number:
Item and revision number of instructions for use:
Status of operation:
Active
Inactive
Name of vendor:
Warranty:
Start: ________________________________________________
End: ________________________________________________
Model: ________________________ SN: ________________________
Instructions on use to starting the cabinet:
1. Training of the responsible party Date: By:
2. Operational test of the cabinet Date: By:
3. Responsible party Tel:
Instructions to users:
The responsible party is trained in use of the cabinet in reference to the user manual
General use of cabinet Objections to the mentioned:
Service & maintenance
The cabinet was delivered without defects/damage.
The cabinet started as specied in the user manual
Set values:
Setpoint temperature ºC
Local alarm settings
High temperature alarm ºC
Low temperature alarm ºC
External alarm settings
(See voltage free contact in user manual)
High temperature alarm ºC
Low temperature alarm ºC
Model /
Setpoint temperature
LHL LLL EHL ELL
RF -20 °C +25 °C -35 °C +25 °C -35 °C
RR +5 °C +25 °C 0 °C +25 °C 0 °C
Factory settings:
Name of trained user: Name of instructor:Signature: Signature:Date:
Model: ________________________ SN: ________________________