75
b i o l ine
74
75
BioCompact II RF610HBioCompact II RR610H
Technical specications Data
Temperature range +2/+20 °C
Ambient temperature range +10/+38 °C
Software variant K6+
K-Value 0.46 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T6 Gc
ATEX certicate –
Gross volume 583 litres
Net volume 419 litres
Dimensions – W x D x H 695 x 875 x 1874/2075 mm
Refrigerant R134a
Refrigerant charge 230 g
Refrigeration capacity at -10 °C 314 Watt
GWP – CO2e N/A
Energy consumption –
default setpoint
2.02 kWh/24h
Heat emission 100 % 334 Watt
Heat emission default setpoint 269 Watt
Nominal consumption 329 Watt
Start current 11.4 A
Sound level 44.2 dB(A)
BioCompact II RR610H
Glass door
Technical specications Data
Temperature range +2/+20 °C
Ambient temperature range +10/+43 °C
Software variant K2+
K-Value 0.31 W/(m2*K)
ATEX marking II 3G Ex nA nC nL llB T6 Gc
ATEX certicate –
Gross volume 583 litres
Net volume 419 litres
Dimensions – W x D x H 695 x 875 x 1874/2075 mm
Refrigerant R134a
Refrigerant charge 230 g
Refrigeration capacity at -10 °C 314 Watt
GWP – CO2e 328.9
Energy consumption –
default setpoint
1.36 kWh/24h
Heat emission 100 % 257 Watt
Heat emission default setpoint 56 Watt
Nominal consumption 314 Watt
Start current 11.4 A
Sound level 44.2 dB(A)
BioCompact II RR610H
Solid door
Technical specications Data
Connection 230 VAC, 50 Hz
Control Unit Gram Control Unit with voltage free contact, alarms and oset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Alarm ports Voltage free contact
Access port 1 pcs. ø24.5 mm
Door Right or left hinged
Material interior PS lining with wall rails in stainless steel
Material exterior White lacquered steel or stainless steel nish
Wall thickness 60 mm
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water
IP class IP21
BioCompact II 610
General data: