Initial setup steps
Installation
Cabinet
Minimum ambient
operating temperature
Maximum ambient
operating temperature
BioCompact II 210, 310, 410, 210/210, 310/210
RR with solid door +10 ºC +35 °C
RR with glass door +10 °C +32 °C
RF +10 °C +35 °C
BioCompact II 610
RR with solid door +10 °C +43 °C
RR with glass door +10 °C +38 °C
RF +10 °C +43 °C
WARNING –
Potential electrostatic hazard
Removing protective packaging and
lm may cause electrostatic discharge.
Protective packaging and lm shall not
be removed in ATEX zones.
I-1*: Due to safety and operating considerations, the cabinet must not be used outdoors.
I-2*: The cabinet should be installed in a dry and suciently ventilated area.
I-3*: To ensure ecient operation, the cabinet should not be installed in direct sunlight or close to heat sources.
The cabinet interior must not be exposed to corrosive atmospheres.
I-6*: The cabinet is shipped with a protective lm that should be removed prior to use.
I-5*: Avoid placement of the cabinet in a chloric/acidic environment due to risk of corrosion.
I-7*: Clean the cabinet with a mild soap solution prior to use.
I-8*: The cabinet is only allowed to lay down for very short durations (for instance handling through a
doorway). If the cabinet has been laying down, the cabinet must stand up-right for at least 24 hours
prior to use. This enables oil in the compressors to run back into place.
I-4*: Ambient operating temperature range
*): Paragraph designation (I-1, I-2 etc) refers to the IQ (Installation Qualication) *): Paragraph designation (I-1, I-2 etc) refers to the IQ (Installation Qualication)