Door gaskets are an important part of a cabinet. Impaired door gaskets can lead to increased humidity, iced up evaporator (and thus
reduced cooling capacity), and in some cases, decreased longevity of the cabinet.
It is therefore very important to be aware of the door gaskets condition. Regular inspection is recommended.
The door gasket should be cleaned regularly with a mild soap solution. If a gasket is to be replaced, please contact your local Gram
BioLine distributor. The illustration below shows the location of the door gasket on a BioCompact II 410.
Cleaning Door gasket
Regular maintenance
The cabinet must not be connected to a power source while cleaning.
The cabinet should be cleaned internally with a mild soap solution (max. 85 °C) at suitable intervals and checked
thoroughly before it is put into operation again.
The compressor compartment and in particular the condenser must be kept free from dust and dirt. This is best done
with a vacuum cleaner and a brush.
It is recommended that the re-evaporation tray is checked regularly for foreign objects and cleaned accordingly.
Do not ush compressor compartment or evaporator with water.
Cleaning agents containing chlorine or compounds of chlorine as well as other corrosive agents, may not be used, as they
may cause corrosion.
The location of the condensers for both bottom- and top-mounted compressors is illustrated below.
Door gasket