102 103
b i o l ine
Measurement – Prerequisites
ID Description Accepted
Yes No
P-1 The cabinet must be empty while conducting tests,
ie without interior ttings
such as drawers, shelves etc.
Attachment:
Notes:
P-2 The measurements must be conducted in accordance to IEC 60068-3-5.
Attachment:
Notes:
P-3 The positioning of the sensors in the cabinet must be documented with a
sketch and/or a photograph.
Attachment:
Notes:
Deviations from the specications dictated in the PQ, are to be reported in the deviation report.
The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectied or
accepted.
Conducted by:
Inspected / veried by:
Name: Signature: Approved
(Yes / No):
Date:
Conducted by:
Inspected / veried by:
Name: Signature: Approved
(Yes / No):
Date:
Measurement – Prerequisites
ID Description Accepted
Yes No
P-4 Measurements made during the PQ tests must be documented and
attached to the PQ.
Attachment:
Notes:
P-5 Specify setpoint temperature: °C
Specify the ambient temperature: °C
Attachment:
Notes:
P-6 Allowed tolerances –
Select the tolerance, according to the model being tested.
Find model-specic tolerances in appendix.
Tolerance: +/- K
Attachment:
Notes:
Deviations from the specications dictated in the PQ, are to be reported in the deviation report.
The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectied or accepted.
Model: ________________________ SN: ________________________ Model: ________________________ SN: ________________________