3-18
Settings and Measurement
L: An alarm is generated when the signal input falls to (or falls below)
the specified alarm level.
CH.1
Alarm generated
Win In: Specifies the upper and lower alarm limits for each channel.
When the signal level goes within (or is within) both limits, an
alarm is generated.
CH.1
Alarm generated
Win Out: Specifies the upper and lower alarm limits for each channel.
When the signal level goes outside (or is outside) both limits,
an alarm is generated.
CH.1
Alarm generated
Lower - Level- Upper .......... Specifies the alarm level(s) for the conditions set in Mode.
L, H: Input a numeric value.
Win In, Win Out: Input numeric values for the upper and lower levels.
Output ................................. Specifies the output channels for alarm generation (4 channels).
Pulse
Sets the alarms for pulse input settings. These conditions can be set when they have been enabled
in the AMP settings.
Mode................................... Sets the same conditions as for "Condition Settings". Please refer to
"Condition Settings" in the previous section.
Lower - Level- Upper .......... Specifies the alarm level(s) for the conditions set in Mode.
L, H: Input a numeric value