2−6
CHAPTER 2 Checks and Preparation
2.5 Connecting the Signal Input Cables
This section describes how to connect the signal input cables.
Terminal Configuration and Signal Types
Shunt resister
Ex: The current is converted to the voltage in
the shunt register.
For 4 to 20mA current input, installing 250
ohms (0.1%) resister for converting 1 to 5V.
* 250 ohms shunt register is the option B551.
Direct current
+ ......................................High-voltage terminal (terminal for high-voltage input signals)
– ......................................Low-voltage terminal (terminal for low-voltage input signals)
b ......................................Dedicated terminal when connecting resistance temperature detector
* Resistance temperature detector input terminals A (+) and B (-) are isolated within each channel. Terminal b is shorted
within all channels.
Item Description
Input configuration Isolated input, scanning
Analog voltage 20, 50, 100, 200, 500 mV/F.S.; 1, 2, 5, 10, 20, 50 V/F.S.; 1-5V
Thermocouples K, J, E, T, R, S, B, N, W (WRe 5-26)
Resistance temperature detector
PT100, JPT100, PT1000 (IEC751)
A/D resolution 16-bit (Effective resolution: About 1/40,000 of the +/- range)
Filter Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the set sampling count is used.
If the sample interval exceeds 30 seconds, the average value of data ob-
tained in a sub-sample (30 seconds) is used.