Analog settings
Specify the conditions for analog signals.
When you use CH ALL to set an input, range and filter, all channels are set to the same values if the input is the same. Range is
set only for the same input channels. However, the range of a channel is not changed if its EU (scaling) is set to On.
Span All Settings is set only for the same range channels.
*If the first channel in a channel group (CH1 if CHs 1 to 10 are displayed) has an input that is set to Off, the input of CH ALL is set to Off.
(1)-1 Input
Selects input condition
Selection item Description
Off Input signal measurement is disabled. No waveform or digital
value is displayed.
Voltage Used for measuring direct-current voltage.
Temperature Used for measuring temperature.
Humidity Used for measuring humidity with the humidity sensor B-530.
In this case, the voltage range will become 1 V, and the EU
settings will not be available.
(1)-2 Range
Selects the range of measurement.
Input item Description
Voltage 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 1-5 V
Temperature TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N, TC-W,
Pt100, JPt100, Pt1000
Humidity No selection available
Available SPAN Settings
<Voltage Ranges>
Range Maximum SPAN Minimum
SPAN
Minimum
Resolution
20mV -22.000 to +22.000mV 0.200mV 0.001mV
50mV -55.00 to +55.00mV 0.50mV 0.01mV
100mV -110.00 to +110.00mV 1.00mV 0.01mV
200mV -220.00 to +220.00mV 2.00mV 0.01mV
500mV -550.0 to +550.0mV 5.0mV 0.1mV
1V -1.1000 to +1.1000V 0.0100V 0.0001V
2V -2.2000 to +2.2000V 0.0200V 0.0001V
5V -5.500 to +5.500V 0.050V 0.001V
10V -11.000 to +11.000V 0.100V 0.001V
20V -22.000 to +22.000V 0.200V 0.001V
50V -55.00 to +55.00V 0.50V 0.01V
1-5V -5.500 to +5.500V 0.050V 0.001V