3-22
Settings and Measurement
Start side source settings ... Specifies conditions to start data capture.
• Off: Starts capturing data unconditionally.
• Level: Starts capturing data when a specified level is reached.
• Alarm: Starts capturing data when the alarm with the specified
number is generated.
• External Input: Starts capturing data when an input signal is received
from an external trigger terminal.
• Date: Starts capturing data when a specified time arrives
(when repeated capturing is set to Off)
: Starts capturing data when a specified time arrives
(when repeated capturing is set to On)
This setting is used when you want to start capturing data at the
same time every day.
Stop side source settings.... Specifies conditions to stop data capture.
• Off: Does not stop data capture by a trigger.
• Level: Stops data capture when the specified level is reached.
• Alarm: Stops capturing data when the alarm with the specified
number is generated.
• External Input: Stops capturing data when an input signal is received
from an external trigger terminal.
• Date: Stops capturing data when a specified time arrives
(when repeated capturing is set to Off).
: Stops capturing data when a specified time arrives
(when repeated capturing is set to On).
This setting is used when you want to stop capturing data at the
same time every day.
• Time: Stops capturing data at a specified time after starting data
capture.
Repeated capturing ............ After a stop side trigger is generated, the next data capture process
begins.
• Off: Does not repeat data capture.
• On: Repeats data capture.
Alarm level settings............. When the setting level conditions are met, the alarm output terminl
outputs an alarm.
The Digital screen showing the channel for which the alarm has been
generated is displayed in red.
Alarm Hold .......................... Specifies whether or not to maintain the alarm status when an alarm is
generated and then canceled.
• Not Maintained: Alarm status is canceled when the alarm is
canceled.
• Maintained: Alarm status is not canceled even though the alarm is
canceled.
Send burnout Alarm ............ When burnout (see on page 3-28) occurs, an alarm signal is output to
the alarm output terminal.
• Does not occur: Alarm signal is not output to the alarm output
terminal when burnout occurs.
• Occurs: Alarm signal is output to the alarm output terminal when
burnout occurs.
Alarm is output to the output terminal set in the Alarm level settings.