Warnings, Cautions, and Errors
In this manual the following two labels identify potentially dangerous or
destructive conditions and procedures.
The WARNING label identifies conditions or practices
that may present danger to the patient and/or user.
The CAUTION label identifies conditions or practices
that could result in damage to the equipment
NOTE: Notes help you identify areas of possible confusion and avoid
potential problems during system operation.
The Corti Otoacoustic Emission Test System should be
charged using only the provided power supply. Injury to
personnel or damage to equipment can result when a
three-prong to two-prong adaptor is connected between
the Corti power supply and an AC outlet.
No modifications of the equipment are allowed by anyone
other than a qualified GSI representative. Modification
of the equipment could be hazardous.
The Corti product has been verified by an independent
laboratory to conform to international standards for EMC
(electromagnetic emissions and immunity). The user is
advised to avoid installation and use of this instrument in
proximity with other devices or equipment that may emit
or be susceptible to electromagnetic interference,
including mobile phones. If the instrument is used
adjacent to other devices or equipment, the user is
instructed to verify that no disturbance is found in the
operation of this or other equipment in proximity.
This icon indicates that patient applied parts of the
instrument conform to IEC 60601-1:2005, Type B
requirements.