EasyManua.ls Logo

GSi CORTITM - Configurations and Test Protocols; Diagnostic Configurations and Protocols

GSi CORTITM
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix E: Configurations and Test Protocols
GSI Corti Diagnostic Configurations
Configuration
Protocols Included
DP Diagnostic:
DP 4s (default - not customizable)
DP 2.0-5.0
DP 1.5-6.0
DP 1.6-8.0
DP 1.5-12
TE Diagnostic:
TE 64s (default - not customizable)
TE 1.5-4.0
TE 0.7-4.0
DP and TE
Diagnostic Combo:
DP 4s (default - not customizable)
DP 2.0-5.0
DP 1.5-6.0
DP 1.6-8.0
DP 1.5-12
TE 64s (default - not customizable)
TE 1.5-4.0
TE 0.7-4.0
DPOAE Diagnostic Default Protocols
Protocol
Name
# of
Frequ
encies
F2 Frequency
L1/L2
Averaging
Time
Pass
SNR
# Passing
Freq. for
Test Pass
DP 4s
4
2.0, 3.0, 4.0, 5.0 kHz
65/55
4 sec
6 dB
3
DP 2.0-5.0
4
2.0, 3.0, 4.0, 5.0 kHz
65/55
4 sec
6 dB
3
DP 1.5-6.0
6
1.5, 2.0, 3.0, 4.0, 5.0, 6.0
kHz
65/55
4 sec.
-
0
DP 1.6-8.0
12
1.6, 2.0, 2.5, 3.2, 3.6, 4.0,
4.5, 5.0, 5.6, 6.3, 7.1, 8.0
kHz
65/55
4 sec.
-
0
DP 1.5-12
12
1.5,2,3,4,5,6,7,8,9,10,11,12
kHz
65/55
4 sec.
-
0
TEOAE Diagnostic Default Protocols
Protocol
Name
# of
Frequency
Bands
Frequency
Center Bands
Averaging
Time (max)
Pass SNR
# Passing
Freq. for
Test Pass
TE 64s
6
1.5, 2, 2.5, 3, 3.5,
4 kHz
64 sec
4 dB
3
TE 1.5-4.0
6
1.5, 2, 2.5, 3, 3.5,
4 kHz
64 sec
4 dB
3
TE 0.7-4.0
6
0.7, 1.0, 1.4, 2.0,
2.8, 4.0 kHz
64 sec.
-
0

Table of Contents