37
CoboSafe-CBSF 325-2810-012US15
MAINTAINING FUNCTIONALITY
13 Maintaining Functionality
The CoboSafe-CBSF force measuring devices should be inspected before and after measure-
ment to ensure proper functioning. Look for apparent damage and excessive pollution. Never
use a damaged device.
If there is any uncertainty regarding proper functioning or precision, e.g., because the device
has been dropped or shows damage, the CoboSafe-CBSF should be sent to the manufacturer
to be inspected.
13.1 Maintenance Schedule
Inspection Interval Maintenance Tasks Personnel
Before each measurement Inspection and cleaning Laboratory scientist,
System integrator
After each measurement Inspection and cleaning Laboratory scientist,
System integrator
After approx. 20 operating
hours or when indicated
Charging batteries Laboratory scientist,
System integrator
Annually Calibration of
CoboSafe-CBSF
Manufacturer or certified ser-
vice personnel
Replacement of dam-
ping elements K1
Laboratory scientist,
System integrator
Every 2 years
Replacing batteries
Manufacturer
After updating standards or
in case of wear
If necessary, replace dam-
ping elements K1, firmware
update
Manufacturer
13.2 Checking and Cleaning
Instructions for checking and cleaning CoboSafe-CBSF force sensors and K1 damping ele-
ments.
Personnel Personal Protective Equipment Materials
■ Laboratory scientist
■ System integrator
■ Protective clothing
■ Protective gloves
■ Safety footwear
■ Soft, lint-free cloth