TRIGger:AUTO
Turns Trigger Auto mode on/off.
Parameters: ON | OFF
Example: TRIG:AUTO OFF
Turns the Trigger Auto mode off.
TRIGger:AUTO?
Returns the Trigger Auto mode.
Return parameter: 0|1, 0=OFF, 1=ON
SAMPle:COUNt
Sets the number of samples.
Parameter: <NR1>(1 ~ 9999) | MIN | MAX
Example: SAMP:COUN 10
Sets the number of samples to 10.
SAMPle:COUNt?
Returns the number of samples.
Parameter: None | MIN | MAX
TRIGger:COUNt
Sets the number of trigger counts.
Parameter: <NR1>(1 ~ 9999) | MIN | MAX
Example: TRIG:COUN 10
Sets the number of trigger counts to 10.
TRIGger:COUNt?
Returns the number of trigger counts.
Parameter: None | MIN | MAX