GDM-834X Series User Manual
96
Save to USB
The USB STO option allows the GDM-8342 to
store each measurement reading to a USB stick.
The USB Store function varies according to
whether the operator mode is set to Simple or
Advance.
When the GDM-8342 starts to save records to
USB, all buttons except for the SHIFT, MENU, 2ND
and left and right arrow keys will be locked and
disabled. Remote control will also be disabled; the
GDM-8342 will stop receiving or transmitting any
commands after it starts to save records to USB.
Save to USB (Simple Mode)
The procedure below describes the save
operation when the Mode is set to Simple.
1. Insert a USB stick into the USB Host port on the
front panel.
2. If the USB stick is recognized by the DMM, the
USB STO icon will be lit. This indicates that the
DMM is ready to save files to the USB stick.
3. Press SHIFT → MENU.
The USB STO icon will flash slowly, indicating
the DMM is saving to USB.
4. To stop saving to USB, press SHIFT → MENU
again.
When the save operation has stopped, the USB
STO icon will stop flashing and will remain lit.