GDS-1000A-U Series User Manual 
128 
MAINTENANCE 
Two types of maintenance operations are 
available: calibrating the vertical resolution, and 
compensating the probe. Run these operations 
when using the oscilloscope in a new environment. 
Vertical Resolution Calibration 
 
1.  Press the Utility key. 
2.  Press the More key twice. 
4.  Press Vertical. The message 
“Set CAL to CH1, then press 
F5” appears at the bottom of 
the display. 
5.  Connect the calibration 
signal between the rear 
panel CAL out terminal and 
the Channel1 input. 
6.  Press F5. The calibration automatically starts.