GW Instek GDS-1000A-U Series - The Accuracy Does Not Match the Specifications; The Oscilloscope will Not Allow a 2 M Waveform to be Saved

GW Instek GDS-1000A-U Series
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FAQ
133
The accuracy does not match the specifications.
Make sure the device is powered on for at least 30 minutes, within
+20°C~+30°C. This is necessary to stabilize the unit to match the
specification.
The oscilloscope will not allow a 2M waveform to be saved.
Make sure that only 1 channel is active. Make sure that the signal
has been triggered and that the STOP or Single key has been pressed.
Ensure the time base is slower than 10 ns/div. See page 106.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com / marketing@goodwill.com.tw.