7.  The Channel1 calibration 
will complete in less than 5 
minutes. 
8.  When finished, connect the 
calibration signal to the 
Channel 2 input and repeat 
the procedure. 
9.  When the calibration is complete the display 
will go back to the previous state. 
1.  Connect the probe between the Channel1 input 
and the probe compensation output (2Vp-p, 
1kHz square wave) on the front panel. Set the 
probe voltage attenuation to x10. 
VOLTS/DIV VOLTS/DIV TIME/DIV
CH 1 MATH CH 2 MENU MENU
Acquire
Display
Utility Help
Run/Stop
VARIABLE
FORCE
Autoset
Cursor
SINGLE
Hardcopy
Measure
Save/Recall
LEVEL
VERTICAL HORIZONTAL TRIGGER
CH1
CAT300V
M
W
15pF
MAX. 300Vpk
1
CH2 EXT  TRIG
CAT300V
M
W
15pF
MAX. 300Vpk
1
X Y
2.  Press the Utility key.