GDS-3000 Series User Manual
126
Trigger Parameter Overview
All the following parameters are common for all
the trigger types unless stated otherwise.
Trigger Source
CH1 ~ 4
Channel 1 ~ 4 input signals
EXT
External trigger input
signal
EXT TRIG
AC Line
AC mains signal
Alternate
Alternate between channel sources for
the trigger source.
EXT Probe
Probe trigger source. Set the probe as
either current or voltage.
Trigger Mode
Auto
(un-
triggered
roll)
The GDS-3000 generates an internal
trigger if there is no trigger event, to
make sure waveforms are constantly
updated regardless of trigger events.
Select this mode especially when
viewing rolling waveforms at slower
timebases.
Normal
The GDS-3000 acquires a waveform
only when a trigger event occurs.
Single
The GDS-3000 acquires a
waveform once when a
trigger event occurs, then
stops acquiring. Press the
Single key to acquire a
waveform again.
Single

Table of Contents