EasyManua.ls Logo

GW Instek GDS-3000 Series - Using Pulse Width Trigger

GW Instek GDS-3000 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CONFIGURATION
135
5. Use the side menu to select the
pulse width trigger source.
Range
Channel 1 ~ 4 (Alternate On/Off),
Line, EXT (Ext Probe: Volt/Current,
Attenuation: 1mX~1kX ), AC Line
6. Press Polarity to toggle the polarity
type.
Range
Positive (high to low transition)
Negative (low to high transition)
7. Press When from the lower bezel.
8. Then use the side menu to select
the pulse width condition and
width.
Condition
> , < , = , ≠
Width
4ns ~ 10s
9. Press Threshold from the lower
bezel to edit the pulse width
threshold.
10. Use the side menu to set the
threshold.
Range
-XXV~XXV
Set to TTL 1.4V
Set to ECL -1.3V
Set to 50%