EasyManua.ls Logo

GW Instek GDS-3000 Series - Page 91

GW Instek GDS-3000 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MEASUREMENT
91
Maximum and
Minimum
1. To set a maximum and minimum
boundary, press Minimum Position
from the bottom menu and use the
Variable knob to select the absolute
minimum position.
Position
Voltage division range
2. Press Minimum Position again to
toggle the menu to Maximum
Position. Choose the maximum
position.
Position
Voltage division range
Note
If a maximum and minimum
boundary is set, the tolerance will
be disabled.
Save Boundary
Limits
1. Select a Maximum Position or a
Tolerance from the lower menu.
2. To save the Go-NoGo test
parameters, press Save Operation.
The Maximum position will be
saved to R1. The tolerance
waveforms will be saved to R1 and
R2.
3. If a Maximum Position was saved,
repeat the save procedure for
Minimum Position. The Minimum
position waveform will be saved to
R2.

Table of Contents

Related product manuals