BASIC OPERATION
39
Using the Marker Function
Description
This section will describe how to activate and
move a normal marker. The noise marker
function will also be used to show how to make
a basic marker measurement.
Operation
1. Use the procedure described in the previous
section to display a signal from the REF out
terminal.
2. Press
Marker
> Select Marker[F1] and select
marker number 1.
3. Press [F2] and turn the marker 1 on.
4. Press Normal[F3] and set the marker position to
12 MHz using either the keypad, scroll wheel or
arrow keys.
5. Press Function[F5]>Marker Noise[F2] and turn
the marker noise function on. The noise marker
function calculates the average noise level over
a bandwidth of 1Hz, referenced from the
marker position.