EasyManua.ls Logo

GW Instek GSP-9300 - Page 54

GW Instek GSP-9300
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GSP-9300 Quick Start Guide
52
Spurious Response
Second Harmonic
Intercept
Preamp off; signal input -30dBm;
0 dB attenuation
+35 dBm
Typical
; 10 MHz < fc < 775 MHz
+60 dBm
Typical
; 775 MHz ≤ fc < 1.5 GHz
Third-order
Intercept
Preamp off; signal input -30dBm;
0 dB attenuation
> 1dBm
300 MHz to 3 GHz
Input Related
Spurious
< -60 dBc
Input signal level -30 dBm, Att.
Mode, Att=0dB; 20-30ºC
Residual
Response
(inherent)
<-90 dBm
Input terminated; 0 dB
attenuation; Preamp off
Sweep
Sweep Time
Range
310 us to 1000 s
Span > 0 Hz
50 us to 1000 s
Span = 0 Hz; Min
Resolution = 10 us
Sweep Mode
Continuous; Single
Trigger Source
Free run; Video; External
Trigger Slope
Positive or negative edge
RF Preamplifier
Frequency Range
1 MHz to 3 GHz
Gain
18 dB
Nominal
(installed as standard)
Front Panel Input/Output
RF Input
Connector Type
N-type female
Impedance
50 ohm
Nominal
VSWR
<1.6 :1
300 kHz to 3 GHz; Input
attenuator ≥ 10 dB
Power for Option
Connector Type
SMB male
Voltage/Current
DC +7V / 500 mA max
With short-circuit
protection

Other manuals for GW Instek GSP-9300