APPENDIX
55
Tracking Generator
5
(Optional)
Frequency Range
100 kHz to 3 GHz
Output Power
-50 dBm to 0 dBm in 0.5 dB steps
Absolute Accuracy
± 0.5 dB
@160 MHz, -10 dBm,
Source attenuation 10
dB, 20 to 30°C
Output Flatness
Referenced to 160 MHz, -10 dBm
100 kHz to 2 GHz
± 1.5 dB
2 GHz to 3 GHz
± 2 dB
Output Level
Switching
Uncertainty
± 0.8 dB
Referenced to -10 dBm
Harmonics
< -30 dBc
Typical, output level = -
10 dBm
Reverse Power
+30 dBm max.
Connector type
N-type female
Impedance
50 ohm
Nominal
Output VSWR
< 1.6:1
300 kHz to 3 GHz,
source attenuation ≥
12 dB
[5] The minimum RBW filter is 10kHz when the TG output is ON.
USB Power Sensor (Optional)
Type
Average power sensor
Model: PWS-06
Interface to Meter
USB cable to GSP9300 Front-Panel USB Host
Connector Type
N-type male, 50 ohm nominal
Input VSWR
1.1: 1
Typical
1.3: 1
Max
Input Frequency
1 to 6200 MHz
Sensing Level
-32 to +20 dBm
Max. Input
Damage Power
≤ 27 dBm