Command Reference
102
Query Response
<function>
Example
SEND> FUNC?<NL>
RET> Cp-D<NL>
11.7.2 FUNCtion:IMPedance:AUTO
The FUNCtion:IMPedance:AUTO command sets the impedance’s LCZ
Automatic selection.
Command
Syntax
FUNC:IMPedance:AUTO {ON,OFF, 0,1}
Example
SEND> FUNC:IMP:AUTO ON<NL>
Query Syntax
FUNC:IMPedance:AUTO?
Query Response
{on,off}<NL>
Example
SEND> FUNC:IMP:AUTO?<NL>
RET> off<NL>
11.7.3 FUNCtion:IMPedance:RANGe
The FUNCtion:IMPedance:RANGe command sets the impedance’s
measurement range.
Command
Syntax
FUNC:IMPedance:RANGe <0-8,MIN,MAX>
Parameter
Where, <0-8,MIN, MAX> is:
0-8 The range number
MIN =Range 0
MAX =Range 8
Example
SEND> FUNC:IMP:RANG 2<NL>//Set
measurement range to [2] 10k.
Query Syntax
FUNC:IMPedance:RANGe?
Query Response
<0-8><NL>
Example
SEND> FUNC:IMP:RANG?<NL>
RET> 0<NL>
11.7.4 FUNCtion:DCR:RANGe
The FUNCtion:DCR:RANGe command sets the DCR’s measurement range.
Command
Syntax
FUNC:DCR:RANGe <0-8,MIN,MAX>
Parameter
Where, <0-8,MIN, MAX> is:
0-8 The range number
MIN =Range 0
MAX =Range 8
Example
SEND> FUNC:DCR:RANG 2<NL>//Set DCR range
to [2] 10k.
Query Syntax
FUNC:DCR:RANGe?