LCR-6000 Series User Manual
111
Example
SEND> FETC:MON?<NL>
RET> +3.88651e+05,+0.00000e+00<NL>
(0: The monitor 2 is OFF)
11.10.6 FETCh:LIST?
The FETCh:LIST? query sets the latest LIST measurement data of the
primary parameters, secondary parameters and comparator results into the
output buffer. Only applicable when in the [LIST MEAS] page view.
Query Syntax
FETCh:LIST?
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result> // spot no:01-10
Example
SEND> FETC:LIST?<NL>
RET> 01,-2.98524e-
12,+3.27673e+00,L,02,+7.11030e-12,+3.48450e-
01,P,03,+7.11322e-12,+5.14944e-02,H,04,-
1.00000e+20,-1.00000e+20,-,05,-1.00000e+20,-
1.00000e+20,-,06,-1.00000e+20,-1.00000e+20,-,07,-
1.00000e+20,-1.00000e+20,-,08,-1.00000e+20,-
1.00000e+20,-,09,-1.00000e+20,-1.00000e+20,-,10,-
1.00000e+20,-1.00000e+20,-<NL>
(-1.00000e+20: The STEP is OFF)
Query Syntax
FETCh:LIST? <spot no>
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result>
Example
SEND> FETC:LIST? 2<NL>
RET> 02,+7.11030e-12,+3.48450e-01,P<NL>
(-1.00000e+20: The STEP is OFF)