LCR-6000 Series User Manual 
39 
    During the measurement, an “LCR OPEN 
measurement in progress” dialog message is 
shown on the display. 
    When the measurement has finished, 
“Correction finished” is displayed. During the 
measurement, you can press the [ABORT] 
soft-key to abort the OPEN correction. 
Step 7.  After the OPEN measurement has finished, 
the meter will enable OPEN correction 
automatically. You can, however, still choose 
to manually choose to enable or disable OPEN 
correction by depressing the ON or OFF soft-
keys. 
4.2.2 Short Correction [SHORT] 
The short correction feature of the LCR-6000 Series 
compensates for any residual impedances (R,X) that may 
exist outside of calibration, such as the impedance of the 
cables and the DUT connection points. 
See Figure 4-4  Residual Impedances. 
Figure 4-4  Residual Impedances 
To perform short correction 
Step 1.  Press the [Measure] key. 
Step 2.  Press the [OPEN SHORT] soft-key. 
Step 3.  Use the cursor key to select the [SHORT] field. 
Enables short correction. 
Disables short correction. 
Starts short correction (AC). 
Starts short correction (DC).