EasyManua.ls Logo

GW Instek LCR-6002 - OPEN;SHORT;SPOT Correction Procedures

GW Instek LCR-6002
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCR-6000 Series User Manual
33
4.1.6 Measurement Speed [SPEED]
SLOW, MED and FAST can be selected for LCR-6000
Series.
SLOW mode will result in more stable and accurate
measurement results.
Procedure for setting measurement speed mode
Step 1. Press the [Measure] key.
Step 2. Use the cursor key to select the [SPEED] field.
Step 3. Use the soft-keys to set measurement speed.
Soft-key
Function
FAST
40 times/sec
*1
MED
10 times/sec
*1
SLOW
3 times/sec
*1
*1. Measurement speed in 300kHz, typical. The
Measurement Speed time is the trigger to the
end of measurement (EOM) output on the
Handler.
Display: [BIN MEAS] Page
[RANGE]: HOLD RANGE
[AVG]: 1
[BIAS]: OFF
[AUTO LCZ]: OFF
[MON 1] [MON 2]: OFF
[DELAY]: 0 ms
[ALC]: OFF
Measurement speed [ms]
Speed
Test Frequency(Hz)
SLOW
MED
FAST
10
1600
1600
1600
20
800
800
800
100
483
160
160

Table of Contents