LCR-6000 Series User Manual
115
COMParator:TOLerance:BIN? <n>
Where,<n> is:
NR1 (1 to 9): Bin number
<NR3:low limit>,<NR3:high limit>
SEND> COMP:TOL:BIN? 2<NL>
RET> 1.00000e-06,2.00000E-6<NL>
11.11.7 COMParator:SLIM
The COMParator:SLIM or COMParator:secondary command sets the
LOW/HIGH limit values for the secondary parameter.
COMParator:SLIM <low value>,<high value>
COMParator:secondary <low value>,<high value>
Where, <low value>,<high value> is:
<low value> NR1,NR2 or NR3: low limit value
<high value> NR1,NR2 or NR3: high limit value
A suffix multiplier can be used with this command.
SEND> COMP:SLIM 0.0001,0.0010<NL>
COMParator:SLIM?
COMParator:secondary?
<NR3:low limit>,<NR3:high limit>
SEND> COMP:SLIM?<NL>
RET> 1.00000e-04,1.00000e-03<NL>
11.11.8 COMParator:BEEP
The :COMParator:BEEP command sets beep mode of the comparator
function.
COMParator:BEEP {OFF,PASS,FAIL}
Where,
OFF Turns the beeper off.
PASS Sounds a beep when the test is passed
(BIN1~BIN9).
FAIL Sounds a beep when the test is failed (OUT).
SEND> COMP:BEEP?<NL>
RET> OFF<NL>