3.1 Product components
Make sure that all components have been received. Refer to Figure 2. If any items are missing or
damaged, contact the manufacturer or a sales representative immediately.
Figure 2 Product components
1 Analyzer 3 Standard solution, 5.0 mg/L as
F-
5 Installation, maintenance and
half-cell electrode kits
2 Standard solution, 0.5 mg/L as
F-
4 Reagent solution, TISAB
3.2 Remove the shipping spacer wire
1. Loosen the screws that secure the flow cell assembly. Remove the shipping spacer wire. Refer to
Figure 3.
2. Tighten the screws in small increments and alternate between the screws, so that the flow cell
assembly is moved down evenly. Tighten the screws until the flow cell assembly is seated against
the pump housing. Do not overtighten.