Operation
6. Rest one of the supplied 30 mm User Fill Option membranes
(Cat. No. 17031) on the
M end. Center the membrane as much
as possible on the collector tube's
M end.
7. Take one of the supplied caps and push it over the membrane
resting on the
M end of the collector tube until it cannot be
pushed any further onto the collector tube.
8. The 'skirt' around the sides of the mounted membrane may be
wrinkled and somewhat uneven; this is OK. However, the
mounted membrane should not be torn, especially at its upper
edge. The skirt should not rise to within less than 3 mm from
the end of the tube. If it is torn or very unevenly mounted,
remove the cap and membrane and repeat from step 6,
discarding the old membrane. There is one extra membrane
supplied for every 10 tubes in case this happens.
9. The collector tube is now ready to use with Micro Dist methods.